Fiber Optic Connector Measuring Interferometers
 
  
 
Standards

The DORC® PMS software supports all major measurement methods and techniques, including but not limited to IEC, TIA and Telcordia (Bellcore) standards. The primary purpose of the such organizations, is to ensure that the measurement methods are standardized to the point that consistent results will be achieved, even when using different manufacturers equipment. For single fiber connectors, three main variables have evolved into fitting regions used for the calculation of the endface geometry parameters:

"D" = Fitting region
"E" = Extracting region
"F" = Averaging region

The calculated values of all measured parameters, including radius of curvature, apex offset and fiber height will all change depending on the values of D, E & F that are used. For this reason it is imperative that when re-measuring, or verifying a measurement using another piece of test equipment, the same D, E & F values MUST be used. For your convenience DORC® prints out the D, E & F parameters used on all reports.

The following fitting region information is correct as of September 25th, 2010:

Organization "D" 
Fitting Region
"E" 
Extracting Region
"F"
Averaging Region
IEC (PC & APC) 250 microns 140 microns 50 microns
TIA (Draft) 250 microns 140 microns 50 microns
Bellcore (GR 326) 250 microns 140 microns 50 microns

 

 
 
 

Send mail to webmaster@dorc.com with questions or comments about this web site.
Copyright© 1992-2011, Direct Optical Research Company
DORC® is a Registered Trademark of Direct Optical Research Company
Last modified: January 7, 2011
Tel: (727) 319-9000   Fax: (727) 319-9006 
  sales@dorc.com 
DORC®, 8725 115th Avenue North, Largo, FL  33773  USA