Multi Fiber Reproducibility and Repeatability
| Measured Parameter |
Reproducability - One Sigma |
Repeatability - One Sigma |
| Angle (X and Y) |
0.015 degrees |
0.005 degrees |
| Fiber Height |
15nm |
10nm | Reproducibility based on 50 measurements re-inserting connector between measurements Repeatability based on 100 measurements without disturbing the connector between measurements MT/MPO connectorreproducability and repeatability is somewhat determined by the polish quality and surface geometry
Single Fiber Reproducibility and Repeatability
| Measured Parameter |
Reproducability - One Sigma |
Repeatability - One Sigma |
| Radius of Curvature |
0.5% |
0.25% |
| Apex Offset |
3.0µm |
1.0µm |
| Fiber Height |
2.5nm |
1.5nm | Reproducibility based on 50 measurements re-inserting connector between measurements Repeatability based on 100 measurements without disturbing the connector between measurements
Single Fiber Measurement Range
| Radius of Curvature |
3mm - Flat |
| Apex Offset |
0-500µm (value calculated if outside of captured frame) |
| Fiber Height |
± 6µm |
Other Specifications
| Principle of Operation |
Michelson Interferometer (non contact system) |
| Measurement Speed |
< 45 seconds for MT/MPO connectors. < 3 seconds for single fiber (ex. auto-focus) |
| Field of View |
3.0mm nominal - in the horizontal axis. Entire MT connector in the vertical axis |
| Operating Wavelengths |
470nm to 645nm depending on mode (solid state sources) |
| Power requirements |
100-120V and 210-250V AC 50/60Hz auto sensing (Desktop PC uses voltage switch) |
Due to continued product improvement, specifications are subject to change without notice.
|